There are 2 fuses in the sc200 controller - what is the other fuse for?

Document ID

Document ID TE2090

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
There are 2 fuses in the sc200 controller - what is the other fuse for?
Summary
sc200 fuse information details
Answer
The sc200 is compatible with both 110 VAC and 220 VAC systems so there are two fuses, L1 on the left and L2 on the right. When connected to 120VAC only the L1 fuse is used. When using 240VAC then 120VAC will be on L1 and 120VAC on L2.