My CL17 chlorine analyser is reading correctly, but the pumps are not turning on.

Document ID

Document ID TE2038

Published Date

Published Date 21/01/2019
Question
My CL17 chlorine analyser is reading correctly, but the pumps are not turning on.
Summary
CL17 Not Operating Pumps
Answer
Check to make sure the relays are working correctly by (pressing 918 DIAG on the Older CL17 analyser or Test Relays on the newer CL17 analyser). Then check the set points and make sure that the alarm is set to LOW and not HIGH in the Relay set up.