What can be done if there are still bubbles coming into my instrument after the installation of the bubble trap head regulator assembly, PN 4668000?

Document ID

Document ID TE2020

Published Date

Published Date 12/03/2017
Question
What can be done if there are still bubbles coming into my instrument after the installation of the bubble trap head regulator assembly, PN 4668000?
Summary
Bubble Trap - 1720E, FT660, SS7, CL17
Answer
Improve the bubble trap by creating more surface area for bubbles to adhere to inside of the trap. Place 3 whiffle balls in the bottom of the trap and then 3 regular golf balls on top of them to hold them down. This can significantly improve the effectiveness of the bubble trap to relieve micro bubbles.