If the AluVer® 3 reagent does not completely dissolve will the test still work?

Document ID

Document ID TE1856

Published Date

Published Date 27/05/2017
Question
If the AluVer® 3 reagent does not completely dissolve will the test still work?
Summary
AluVer® 3 reagent must competely dissolve in the sample
Answer
AluVer® 3 Reagent must completely dissolve in the sample no visible black specks can remain. To get the AluVer® 3 to dissolve try heating the sample in a water bath to dissolve.Cool before adding the Bleaching 3 reagent and proceed with the test