If a sample has turbidity is there any pretreatment necessary before using the TNT890 Metals Prep Set?

Document ID

Document ID TE1714

Published Date

Published Date 27/05/2017
Question
If a sample has turbidity is there any pretreatment necessary before using the TNT890 Metals Prep Set?
Summary
TNT890 pretreatment for samples with turbidity
Answer
If the sample exhibits turbidity after digestion, eliminate it by filtering. Use a syringe filter (catalog number 2226100) and a 10-mL syringe (catalog number 2202400).