Can the Troemner weights (catalog number 2617600), which are ASTM Class 1 Certified, be considered equivalent to any other E1, F1, or M1 certifications?

Document ID

Document ID TE1638

Published Date

Published Date 13/04/2020
Question
Can the Troemner weights (catalog number 2617600), which are ASTM Class 1 Certified, be considered equivalent to any other E1, F1, or M1 certifications?
Summary
Troemner weights and E1, F1, M1 certifications
Answer
These weights are not equivalent to E1, F1, or M1 certifications, which are higher class certifications. More information on these certifications can be found on the manufacturers site. These weights (ASTM Class 1) are equivalent to OIML Class E2 however. Sample certificates can be viewed here. Tolerances are different between the classes so they do not have equivalency.