How is the System Verification Chemkey used with the SL1000?

Document ID

Document ID TE1507

Published Date

Published Date 21/01/2019
Question
How is the System Verification Chemkey used with the SL1000?
Summary
Using the Verification Chemkey to validate the SL1000 Portable Parallel Analyser (PPA)
Answer
To use the System verfication chemkey (Product # 9436800) follow the instructions below.
1. Push the Menu button
2. Select Verify Cal
3. Insert the Verification Chemkey in slot 1 and allow instrument to verify slot.
4. Repeat step 3 for slots 2-4.

If this procedure is not followed and the Verification Chemkey is inserted while the meter is in measurement mode, the meter will display an error message, "Verify Only".