The Test Filter Set instructions state to perform multiple programs and a wavelength scan. Is there an easier way to run the filters?

Document ID

Document ID TE1441

Published Date

Published Date 27/05/2017
Question
The Test Filter Set instructions state to perform multiple programs and a wavelength scan. Is there an easier way to run the filters?
Summary
LZV537 Procedures
Answer
Current spectrophotometers have an Optical Check function that can be used with the Test Filter Set (catalog number/order code LZV537). To find this feature, from the main menu select System Checks>Optical Checks>Verfiication set. This menu option will allow input of the lot information for the filters, walk through through the process of checking the instrument, and give a pass/fail message at the end of the readings.