Question
Can a relay be set to trigger upon a warning on the 5500 sc analyzer?
Summary
5500 sc analyzer warning level settings.
Answer
In addition to standard relay setup for control, the relays can also be configured to trigger upon a warning. In order to set the relay, take the following path:
Menu > System Setup > Set Outputs > Relay Setup > Select Relay (A, B, C, D) > Source (select the analyzer name) > Function : Warning > Activation : Warning Level
Listed below are the available warnings to choose from. Any or all can be chosen. Putting a check mark in the box to the left of the warning by pushing the left arrow button will select it to trigger the relay.
Menu > System Setup > Set Outputs > Relay Setup > Select Relay (A, B, C, D) > Source (select the analyzer name) > Function : Warning > Activation : Warning Level
Listed below are the available warnings to choose from. Any or all can be chosen. Putting a check mark in the box to the left of the warning by pushing the left arrow button will select it to trigger the relay.
- High dark current
- Reagent pressure
- Unstable readings
- Low light level
- Reagent low
- Standard solution low
- Sample flow low
- Heater failure
- Fluid leak
- Temperature sensor
- Case temp high
- Failed cal
- Sample 1 missing
- Sample 2 missing
- Sample 3 missing
- Sample 4 missing
- Sample 5 missing
- Sample 6 missing
- Shutdown mode
- Sample pressure > X
- Sample pressure < X
- Carl standard empty
- Auto cal disabled
- Analyzer setup
- Temp out of range
- Open heater circuit
- Reagent pressure high
- Fan failure
- Overfeed timer
- Restart to accept SW
- Network comm loss
- Link2sc job failure