What TOC standard should be used to calibrate the BioTector TOC analyzer?

Document ID

Document ID TE2107

Published Date

Published Date 06/03/2019
Question
What TOC standard should be used to calibrate the BioTector TOC analyzer?
Summary
BioTector TOC standards recommendation.
Answer
You want to use a TOC standard that is about 80% of the range of the BioTector analyzer. The standard can either be prepared according to the directions listed in the user manual, or purchased from Hach. Please contact technical support with the range of your analyzer so that we can provide a Product # for the appropriate TOC standard.
Attachments
Attachments

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.