What do the units NH3-N mean?

Document ID

Document ID TE149

Published Date

Published Date 18/08/2020
Question
What do the units NH3-N mean?
Summary
Explanation of the nitrogen, ammonia units NH3-N chemical form
Answer
Ammonia (NH3) and nitrogen, ammonia (NH3-N) are different expressions of the chemical forms of ammonia. The NH3-N form uses the molecular weight of only the nitrogen atoms. The NH3 form uses the molecular weight of the entire ammonia molecule, 1 nitrogen atom + 3 hydrogen atoms.

The nitrogen-ammonia NH3-N form is common because it relates to other nitrogen compounds such as nitrogen, nitrite NO2-N or nitrogen, nitrate NO3-N.

Provided below is the conversion chart to convert NH3 to NH3-N
NH3 = NH3-N * 1.21589

Assumed atomic weights:
H: 1.008, N: 14.007

Please also see, What is the relationship between Ammonia (NH3) and other nitrogen groups

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.